J4 ›› 2010, Vol. 27 ›› Issue (2): 234-241.

• 激光应用 • 上一篇    下一篇

用MFRSR仪器观测气溶胶光学厚度

王莉萍1,2, 赵凤生3, 李占清4   

  1. 1 中国气象科学研究院, 北京 100081
    2 国家气象中心,北京 100081 
    3 国家卫星气象中心,北京  100081 
    4 马里兰大学,马里兰 21228
  • 出版日期:2010-03-28 发布日期:2010-03-05
  • 基金资助:

    国家自然基金重点课题(40637035)

Observing aerosol optical depth with MFRSR

WANG Li-Ping1,2, ZHAO Feng-Sheng3, LI Zhan-Qing4   

  1. 1 Chinese Academy of Meteorological Sciences,,Beijing 100081,China 
    2 National Meteorological Center,Beijing 100081, China 
    3 National Satellite Meteorological Center,Beijing 100081,  China 
    4 Maryland University, Maryland 21228
  • Published:2010-03-28 Online:2010-03-05

摘要:

气溶胶对激光大气传输有着重要的影响。MFRSR(多光谱旋转遮蔽影带辐射计)是一种用于地基辐射和气溶胶测量的仪器。该仪器使用自动旋转影带技术同时在七个波段交替进行总的水平辐射和漫射水平辐射测量,然后推算得出直接辐射。其中6个波段的中心波长分别是414.3nm,495.3nm,613.7nm,671.5nm,867.6nm,939.3nm,还有一个硅探测器进行宽波段太阳总辐射测量。本文首先介绍了MFRSR仪器及其定标和资料处理方法,然后利用香河观测站2004-2005年MFRSR观测资料,分析了气溶胶的统计特性。为了说明利用MFRSR观测气溶胶光学厚度的可靠性,我们将MFRSR与AERONET的观测结果进行了比较,结果表明二者在500nm、670nm和870nm三个波段的平均偏差分别为-0.021,-0.009,-0.004;标准差分别为0.067,0.051,0.050。文中还对造成二者偏差的原因进行了讨论。

关键词: MFRSR, 定标, 气溶胶光学厚度

Abstract:

MFRSR (multi-filter rotating shadow-band radiometer) is a ground-based instrument and used to make measures of irradiance and aerosol. It uses automated rotating shadow-band technique to alternatively make measurements of total horizontal and diffuse horizontal irradiances at seven wavelengths simultaneously, and then deduce direct normal irradiance. In addition to broadband Silicon detector measures total solar irradiance, center wavelengths of the other six wave bands are 414.3nm, 495.3nm, 613.7nm, 671.5nm, 867.6nm and 939.3nm. This paper introduces the instrument of MFRSR, the method of calibration and data processing, and then analyses statistical properties of aerosol using observation of MFRSR during 2004-2005 in Xianghe. In order to illuminate the reliability of observing aerosol optical depth with MFRSR, we compare the results of MFRSR with AERONET data. It shows the mean differences of the two results respectively are -0.021,-0.009,-0.004 at the wavelengths of 500nm, 670nm, 870nm, and the standard deviations are 0.067,0.051,0.050. The reasons for deviation of the two are also discussed.

Key words: MFRSR, calibration, aerosol optical depth