Calibration of polarimetric Fourier transform infrared spectrometer

GONG Zhongqing1,2, SUN Xiaobing1, ZHANG Qiao1, HONG Jin1   

  1. 1 Remote Sensing Laboratory , Anhui Institute of Optics and Fine Mechanics , Chinese Academy of Sciences , Hefei 230031 , China; 
    2 Artillery Academy of People's Liberation of Army, Hefei 230031, China
  • Published:2011-03-28 Online:2011-03-03

Abstract:

The infrared polarization properties of target can be measured by polarimetric Fourier Transform Infrared(FTIR)Spectrometer. In order to obtain spectral radiance distribution of the target, the instrument must be calibrated in IR waveband. The experimental results indicate that the instrument response functions of the polarimetric FTIR Spectrometer are inconsistent at different polarized orientations. It was caused by the instrumental polarization sensitivity. The polarization characteristics of the instrument were measured and the instrument response function and its offset were acquired. The influence of instrumental polarization sensitivity is eliminated using the result of calibration, Which lay solid foundations for improving polarization detection accuracy of the infrared targets.

Key words: Fourier optics, instrument response functions, calibration, polarization, infrared