Chinese Journal of Quantum Electronics

Previous Articles     Next Articles

Space X-ray CCD driver and processing Circuit Design and Calibration

PENG Jilong   

  1. Beijing Institute of Spacecraft Environment Engineering,Beijing 100094,China
  • Published:2019-03-28 Online:2019-03-20

Abstract: The performance of a CCD, such as readout noise, determines the final data quality of a payload. Commercial CCD sensor cannot be applied in space, neither in X-ray detection. The structure of an X-ray sensing CCD was explained, taking E2V space qualified product for example. Drive circuit met the time and voltage demand was designed with FPGA and an integrated drive circuit. CCD signal amplified and filtered by a gain adjustable CCD signal process electronic. The calibration methods of gain and readout noise using Fe55 radioactive sources are also described. The measured gain of CCD and its support circuit is 5.1 e-/DN, and the readout noise is 37 e-.

Key words: X-ray optics, CCD parameters calibration, driver and processing circuit, gain, readout noise