Chinese Journal of Quantum Electronics ›› 2024, Vol. 41 ›› Issue (3): 455-462.doi: 10.3969/j.issn.1007-5461.2024.03.005

Previous Articles     Next Articles

One‐point calibration LIBS based on Saha‐Boltzmann plot and self‐absorption correction

HU Zhenlin 1#, WANG Tianze 1#, GUO Lianbo 2*, LIN Nan 1*   

  1. ( 1 Shanghai Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Shanghai 201800, China; 2 Wuhan National Laboratory for Optoelectronics, Huazhong University of Science and Technology, Wuhan 430074, China )
  • Received:2023-10-19 Revised:2023-11-13 Published:2024-05-28 Online:2024-05-28

Abstract: To further improve the quantitative analysis accuracy of one-point calibration laser-induced breakdown spectroscopy (OPC-LIBS), the OPC-LIBS based on Saha-Boltzmann plot and self-absorption correction was proposed. Ti, V and Al in two titanium alloy samples were quantified using the method, and the quantitative results were compared with those of OPC-LIBS based on Boltzmann plot, OPC-LIBS based on Saha-Boltzmann plot, and OPC-LIBS based on Boltzmann plot and self-absorption correction. The results show that the proposed method can obtain the smallest quantitative error among the four methods, with mean relative errors (MREs) of 4.11% and 3.99% for the three elements in two titanium alloy samples, respectively, indicating that the proposed method can effectively improve the quantitative analysis accuracy of OPC-LIBS.

Key words: spectroscopy, one-point calibration laser-induced breakdown spectroscopy, Saha-Boltzmann plot, self-absorption correction, plasma temperature

CLC Number: